Features | Technical indicator |
AC withstand voltage test | Rated output capacity | 200VA(5000V/40mA)optional:500VA; |
Output voltage setting | Range/Accuracy | Range:(100~5000)V, Resolution: 1V, ±(1% × setting + 5V) |
Output frequency setting | Range/Accuracy | 50Hz/60 Hz, ±0.1 Hz |
Alarm current setting | Upper limit range /Accuracy | Range:(0.00~100.00)mA, Resolution: 0.01mA, ±(1% × setting + 5 counts) |
Lower limit range/Accuracy | Range:(0.00~9.999)mA, Resolution: 0.001mA, ±(1% × setting + 5 counts) |
Time setting | Range of test time/ Accuracy | Range: 0,(0.5~999.9)s,0 is infinite, Resolution is 0.1s, ±(0.1% × setting + 2 count ) |
|
Range of ramp-up time/ Accuracy | Range: 0,(0.1~999.9)s,0 is off, Resolution is 0.1s, ±(0.1% × setting +2 count) |
Range of ramp-down time/ Accuracy | Range:0,(0.1~999.9)s, 0 is off,resolution is 0.1s, ±(0.1% × setting + 2 count) |
Arc detection | 1~9 (9 corresponds to most sensitive), 0 means the arc function is Off. |
|
DC withstand voltage test | Rated output | 60VA(6000VDC/10mA) |
Output voltage setting | | Range/Accuracy | Range:(100-6000)VDC, Resolution: 1V; ±(1% × setting + 5V) |
Alarming current setting | | Upper limit range/Accuracy | Range:(0.0~10000)μA, Resolution: 0.1μA/1μA, ±(1% × setting + 5 counts) |
| Lower limit range/Accuracy | Range:(0.0~999.9)μA, Resolution: 0.1μA, ±(1% × setting + 5 counts) |
Time setting | | Range of test time/ Accuracy | Range: 0, (0.5~999.9)s, 0 is infinite, Resolution: 0.1s, ±(0.1% × setting + 2 count) |
| Range of ramp-up time/ Accuracy | Range: 0,(0.4~999.9)s, 0 is off, Resolution: 0.1s, ±(0.1% × setting + 2 count) |
| Range of ramp-down time/ Accuracy | Range: 0,(1.0~999.9)s, 0 is off, Resolution: 0.1s, ±(0.1% × setting + 2 count) |
Arc detection | 1~9(9 corresponds to most sensitive), 0 means that arc function is Off. |
Insulation resistance test | Rated output | 2500VDC/50GΩ |
Output voltage | Range/Accuracy | Range:(100-2500)VDC, Resolution: 1V, Accuracy: ±(1% × setting + 5V) |
Alarm resistance setting | Range of upper/lower limits | Range: 0.1MΩ-50000MΩ, upper limit includes no upper limit setting |
Accuracy | 100-499V: 0.10MΩ-999.99MΩ, 1000-2000MΩ, ±(5% × setting + 2 counts) 500-2500V: 0.10-999.99MΩ, ±(2% × setting + 2 counts) 1000-9999MΩ, ±(5% × setting + 2 counts) 10000-50000MΩ, ±(15% × setting + 2 counts) |
|
Time setting | Range of ramp-up time/ Accuracy | Range: 0,(0.1-999.9)s, 0 is infinite,, Resolution: 0.1s, ±(0.1% × setting + 2 count) |
Range of delay time/ Accuracy | Range: 0,(0.5-999.9)s, 0 is off, Resolution: 0.1s, ±(0.1% × setting + 2 count) |
Range of ramp-down time/ Accuracy | Range: 0,(1.0~999.9)s, 0 is off, Resolution: 0.1s, ±(0.1% × setting + 2 count) |
Grounding resistance test | Rated output | Max test current 32A, max resistance 600mΩ, open-circuit voltage <12V |
Output current setting | Range/ Accuracy | Range:(2.0A-32.0A)AC, Resolution: 0.1V, ±(1% × setting + 2 counts) |
Output voltage setting | Range/ Accuracy | Range:(3.0-10.0)VAC, Resolution: 0.1A, ±(1% × setting + 2 counts), in the open circuit case |
Output frequency setting | Range/ Accuracy | 50Hz/60Hz, accuracy: ±0.1Hz |
|
Alarm limit setting | Range of resistance upper/lower limit | 3 A≤ output current ≤10A: 10~600 mΩ 11A≤ output current ≤25A: 10~300 mΩ 26A≤ output current ≤32A: 10~200 mΩ Resolution:0.1/1mΩ |
Accuracy | <100 mΩ, ±(1% × setting + 1 mΩ); ≥100 mΩ, ±(1% × setting + 2 counts) |
Test time setting | Range/ Accuracy | Range: 0,(0.5~999.9)s, 0 is infinite, resolution: 0.1s, ±(0.1% × setting + 2 count) |
Leakage current test | Test method | Single-phase load, leakage current at operating temperature (dynamic) and leakage current at non-operating temperature (static), G-L, G-N, AUTO (G-L, G-N); external isolation transformer or variable frequency power supply is required to provide the required voltage and frequency. |
|
|
|
|
|
|
State of operating power supply | Polarity: on, off, automatic; neutral: on, off; ground: on, off |
Built-in 8 human networks (Standard A/F/H networks, other network optional, choose C/D/E network for testing of medical device)
| |
Test voltage output | Single-phase, external isolation transformer or variable frequency power supply is required to provide the voltage and frequency required for testing. |
Load current | Upper limit: 40A |
Contact current/leakage current upper/lower limit setting (RMS) | Range: 0.0μA~20.00mA, Resolution: 0.1μA/1μA/0.01mA
DC, 15Hz≤f≤100kHz: ±(1.5% × setting + 10 counts) 100kHz<f≤1000kHz: ±(5% × setting) |
Contact current/leakage current upper/lower limit setting (peak) | Range of upper/lower limit: 0.0pA~30.00mA, Resolution: 0.1μA/1μA/0.01mA DC: ±(2% × setting + 2 μA) 15Hz<f≤1000kHz: ±(10% × setting + 2 μA) |
Contact current/leakage current upper/lower limit setting (AC component,optional for testing of medical device) | Range: 0.0μA~20.00mA, 0.0μA~20.00mA, Resolution: 0.1μA/1μA/0.01mA Accuracy: ±(3% × setting + 5 counts) 30Hz<f≤100kHz: ±(2% × setting + 3 counts) 100kHz<f≤1000kHz,10.0μA~999.9μA: ±(5% × setting) 1000μA~7999mA:15Hz≤f≤l00kHz: ±(2%× setting +3 counts) 100kHz<f≤1000kHz,10uA~7999μA: ±(5% × setting) 8.00mA~20.00mA: 15Hz≤f≤1000kHz, 0.01mA~20.00mA: ±(5% × setting) |
Contact current/leakage current upper/lower limit setting (DC component,optional for testing of medical device | Range: 0.0μA~20.00mA, Resolution: 0.1μA/1μA/0.01mA 0.0μA~999.9μA: ±(2% × setting + 3 counts) 1000μA~7999μA: ±(2% × setting +3 counts) 8.00mA~20.00mA: ±(5% × setting) |
Test time | Range: 0, (1~999.9), 0 is infinite, Resolution: 0.1s, ±(0.1% × setting + 2 count ),(when test method is AUTO(G-L,G-N),each gets half time) |
Power parameter measurement | Alarming function | You can choose to set power, voltage or current |
Power upper/lower limit setting(In fact, the upper limit is determined by the power of the power supply) | Range: 0.10W~299.99W, 0.3000kW~1.4999kW, 1.500kW~10.000kW Resolution: 0.01W/0.0001kW/0.001kW PF>0.5: ±(0.1% × reading + 0.1% × measuring range) PF≤0.5: ±(0.4% × reading + 0.1% × measuring range) |
Voltage upper/lower limit setting | When voltage alarm function is enabled Range: 0.00V~300.0V, resolution: 0.01V/0.1V; ±(0.1% × setting + 0.1% × measuring range), 45Hz≤f≤65Hz |
Current upper/lower limit setting (In fact, the upper limit is determined by the power of the power supply) | When current alarm function is enabled Range: 0.010A~0.999A, 1.000A~4.999A, 5.00A-40.00A Resolution: 0.001A, 0.01A ±(0.1% × setting + 0.1% × measuring range),45Hz≤f≤65Hz |
Test time | Range: 0,(0.5~999.9)s, 0 is infinite, Resolution: 0.1s, ±(0.1% × setting +2 count) |
Low voltage starting test | Current upper/lower limit setting | Range: 0.010A~0.999A, 1.000A~4.999A, 5.00A~40.00A Resolution: 0.001A, 0.01A ±(0.1% × setting + 0.1%× measuring range), 45Hz≤f≤65Hz |
Voltage measurement | Range: 10.00-300.00V, peak factor: ≤1.6, Resolution:0.01V/0.1V; ±(0.1% × reading + 0.1% × range), 45Hz≤f≤65Hz |
Current measurement | Range: 0.010A~0.999A, 1.000A~4.999A, 5.00A~40.00A Peak factor: ≤1.6, resolution: 0.001A, 0.01A; ±(0.1% × setting + 0.1% × measuring range), 45Hz≤f≤65Hz |
Test time | Range: 0,(0.5~9.999)s, 0 is infinite, Resolution: 0.1s, ±(0.1% × setting + 2 count) |
LN circuit detection | Detection method | DC constant current power supply |
Resistance upper/lower limit setting | Range:(1.0~999.9)Ω, Resolution: 0.1Ω, ±(2% × reading + 5 counts) |
Test time | Range: 0, (0.5~9.999)s, 0 is infinite, Resolution: 0.1s, ±(0.1%×setting + 2 count) |
Other configuration of equipment | AN1651H(F) | Built-in Ainuo 6kW variable frequency power supply |
AN1640H(F) | External isolated power supply or optional built-in 500W power supply is required |
Dual station test | Optional as required |
Leakage test human simulation network | Standard A//F/H common networks, other networks are optional |
Dimensions | 1640H | 426*177*550(W×H×D,mm) |
1651H | 483*1355*600(W×H×D,mm) |